Chem 155 Unit 12 215 <strong>of</strong> 316IR Spectroscopy is Difficult!• IR beams are invisible to the eye.• IR bands are narrow, so Δλ EFF must be small.• SiO 2 absorbs IR radiation.• Water and other solvents absorb IR radiation.• IR photons are low energy, so they are hard todetect!Why can’t you use a photomultiplier tube in the IR?Infrared radiation has a longer wavelengththan visible radiation:λ IR > λ VIS– so –IR radiation has a lower frequency:ν IR = c/λ IR– so –IR radiation has a lower energy:– so –Photomultiplier tubes don’t work because:E IR = hν IR ,For a PMT to work, the photon energy must be greaterthan the work function <strong>of</strong> the metal!E ≥ φButE IR ≤ φPage 215 <strong>of</strong> 316