INTERLOCK/STATOR<br />

OVERVIEW<br />

This section describes the operation of the following circuits:<br />

• Generator Interlock<br />

• Pre-charge<br />

• Stator<br />

These circuits are initialized and checked by the Technique Processor during the boot<br />

sequence. If a failure is detected during boot-up, the system will hang, not completing the<br />

boot sequence. An error message may be displayed on the Control Panel Display.<br />

After system boot, the Technique Processor continues to monitor these circuits. An error<br />

message is displayed on the Control Panel Display and X-rays are prohibited, if a failure<br />

is detected.<br />

REFERENCE DIAGRAMS<br />

The theory in this section is most easily understood when using the following figures and<br />

schematics. Use the correct schematics for your system.<br />

Generator Controller Assembly Drawing<br />

Drawing # 00-875392 - All serial numbers<br />

Analog Interface PCB<br />

Schematic #00-876738 - Assembly #00-876740 - All system serial numbers<br />

Mainframe Motherboard<br />

Schematic #00-875539 - Assembly #00-900588<br />

System serial numbers 69-0001 thru 69-2000<br />

Schematic #00-878396 - Assembly #00-900970<br />

System serial numbers 69-2001 and up<br />

Power Motor Relay PCB<br />

Figure 4<br />

Figure 5<br />

Schematic #00-875997 - Assembly #875999<br />

All serial numbers