16 Mechanical Assemblies<br />

CONTROL PANEL HOUSING REMOVAL<br />

The Control Panel Housing must be removed before the Wig Wag Assembly can<br />

be accessed.<br />

WARNING... Electrical circuits inside the equipment use voltages that are capable<br />

of causing serious injury or death from electrical shock. Use<br />

appropriate precautions.<br />

1. Turn the Workstation keyswitch to the OFF position.<br />

2. Remove the Hole Plug and 1 Torx screw securing the Cross Arm Brake<br />

Handle.<br />

3. Remove 2 Torx screws from the bottom of the Cross Arm Lower Cover.<br />

4. Lift the Cross Arm Center Cover up carefully and unplug switch wiring on the<br />

Control Panel Processor PCB and the Distribution PCB.<br />

5. Remove 4 Torx screws securing the Cross Arm Lower Cover and lower the<br />

cover so that the Wig Wag Assembly can be accessed.<br />

6. Remove the Hole Plug from the bottom of the Wig Wag brake handle and<br />

loosen the screw.<br />

7. Remove the Wig Wag brake handle.<br />

8. Remove 2 flathead Allen screws in the bottom of the retaining plate and<br />

remove the retaining plate.<br />

9. Loosen the set screw.<br />

10. Turn the actuator screw clockwise using a standard slotted screwdriver until<br />

the brake assembly can be removed.