02.05.2018 Views

GE_OEC_9600_C-Arms

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Image System Calibration 11<br />

CONFIRM MONITOR SIZE & LINEARITY<br />

1. Confirm the left and right monitors are adjusted correctly, referring to Figure 9, by<br />

measuring the following size specification:<br />

• 5" (± 0.125) from the center to the horizontal & vertical edges<br />

NOTE: The monitor CRT curves and is not flat. The measuring viewpoint must<br />

therefore be perpendicular to the monitor screen.<br />

"B"<br />

5"<br />

5"<br />

"A"<br />

Figure 9 - Image Size Measurement<br />

2. Also verify the vertical linearity of the monitors by measuring the largest distance<br />

between the vertical lines (A), and the smallest distance between the vertical lines<br />

(B). The distance should be approximately 1 inch between each line. Calculate the<br />

Linearity Error using the following formula:<br />

A - B X 100% Linearity error must be less than 10%.<br />

A<br />

NOTE: IF THE MONITORS ARE WITHIN SPECIFICATIONS, proceed to the Adjust<br />

Camera Focus portion of this procedure. IF THE MONITORS ARE NOT<br />

WITHIN SPECIFICATIONS, adjust the monitors by continuing to Step 3.