Generator Calibration 9<br />

20. Remove the necessary covers to access the X-ray Regulator PCB and jumper<br />

across E1 on the X-ray Regulator PCB with a shorting plug to disable the mA servo.<br />

21. Plug the system into the AC outlet. Turn on the battery circuit breaker and power up<br />

the system.<br />

22. Make an exposure in the MANUAL FLUORO mode and verify that the DRO is<br />

working.