30 Generator Calibration<br />

B. mA limits cannot exceed previous entries. This is shown as an error in the right<br />

column for entry 95 kV. The 4.9 mA value exceeds the previous 4.8 mA value.<br />

CORRECT<br />

INCORRECT<br />

kVp Max mA kVp Max mA<br />

40 5.0 40 5.0<br />

. . . .<br />

. . . .<br />

90 4.9 90 4.8<br />

. . . .<br />

95 4.4 95 4.9<br />

. . . .<br />

120 2.9 120 2.9<br />

C. A minimum of three entries must be entered. This is shown as an error in the<br />

right column with only two entries at 40 and 120 kV and nothing for the 90 kV<br />

entry.<br />

NOTE:<br />

CORRECT<br />

INCORRECT<br />

kVp Max mA kVp Max mA<br />

40 5.0 40 5.0<br />

. . . .<br />

. . . .<br />

90 4.9 90 .<br />

. . . .<br />

. . . .<br />

. . . .<br />

120 2.9 120 2.9<br />

Even though three entries will allow the software to interpolate the mA Limit File for<br />

the rest of the values, values must be entered at every 5 kV point to ensure better<br />

resolution of the mA limiting.<br />

D. The 120 kV entry must contain a value. This is shown as an error in the right<br />

column.<br />

CORRECT<br />

INCORRECT<br />

kVp Max mA kVp Max mA<br />

40 5.0 40 5.0<br />

. . . .<br />

. . . .<br />

90 4.9 90 4.9<br />

. . . .<br />

. . 95 4.4<br />

. . . .<br />

120 2.9 120 .