02.05.2018 Views

GE_OEC_9600_C-Arms

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

4 Generator Calibration<br />

CALIBRATION TIPS<br />

1. At all times the Dynalyzer digital readout should equal the readout on the left<br />

monitor of the Workstation.<br />

2. During calibration of the .3 mm and .6 mm focal spots, the duty cycle should never<br />

reach 1225.<br />

3. All shots of less than 1 mA are rejected.<br />

4. The system should not take more than eight shots to achieve 1 mA.<br />

5. The 0.3 mm focal spot calibration should follow these guidelines:<br />

A. 40 kV, mA≅45 With each increment of 25 for duty cycle mA<br />

should increment by 2 to 3.<br />

B. 60 kV, mA≅50 With each increment of 25 for duty cycle mA<br />

should increment by 3 to 4.<br />

C. 80 kV, mA≅50 With each increment of 25 for duty cycle mA<br />

should increment by 3.<br />

D. 100 kV, mA≅40 With each increment of 25 for duty cycle mA<br />

should increment by 3.<br />

E. 120 kV, mA≅33.3 With each increment of 25 for duty cycle mA<br />

should increment by 3.<br />

6. The 0.6 mm focal spot calibration should follow these guidelines;<br />

A. 50 kV, mA≅75 With each increment of 25 for duty cycle mA should<br />

increment by 8.<br />

B. 60 kV, mA≅66.7 With each increment of 25 for duty cycle mA should<br />

increment by 8 or 9.<br />

C. 80 kV, mA≅50 With each increment of 25 for duty cycle mA should<br />

increment by 8 or 9.<br />

D. 100 kV, mA≅40 With each increment of 25 for duty cycle mA should<br />

increment by 9 or 10.<br />

7. During the verification run, shots will have a duration of 1500 mS. If the batteries are<br />

weak this is where they will cave in.<br />

8. At the end of the calibration, the duty cycle, calibration coefficients and the<br />

verification data are merged to create the ALL.DAT file.<br />

9. Before sending the generator diskette to <strong>OEC</strong> for archiving make sure that the<br />

following files were created in the following order:

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!