Image System Calibration 15<br />

8. Select the NORMAL, MAG1, and MAG2 field sizes and verify the following<br />

minimum line pairs/mm can be achieved in the field selected.<br />

FIELD SIZE 60 Hz lp/mm 50 Hz lp/mm<br />

NORMAL 1.4 1.2<br />

MAG1 2.0 1.7<br />

MAG2 2.4 2.0<br />

9. You may need to readjust focus to meet the above specifications.