02.05.2018 Views

GE_OEC_9600_C-Arms

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>9600</strong> Imaging Workstation Service Manual xxi<br />

ADJUST CAMERA FOCUS<br />

Follow the procedures in the “Adjust Camera Focus” section in the Beam<br />

Alignment chapter (<strong>9600</strong> System Service Manual), with this exception:<br />

near the end of the procedure, when selecting field sizes to verify the<br />

minimum line pairs/mm, use the following resolutions.<br />

FIELD SIZE 60 Hz lp/mm 50 Hz lp/mm<br />

NORMAL 1.0 1.0<br />

MAG1 1.3 1.4<br />

MAG2 1.6 2.0