MA <strong>GE</strong>NERATION<br />

OVERVIEW<br />

Tube current (mA) is regulated by a dynamic, self correcting servo control loop<br />

comprised of a drive PAL, control amplifier, error amplifier, sense circuit, variable<br />

filament voltage regulator, and a duty-cycle controlled filament drive circuit.<br />

In the Auto mode, the Technique Processor PCB varies the mA according to the video<br />

level signal received from the CCD camera until the proper video level is achieved.<br />

In the Manual mode, the operator varies the mA.<br />

Power for X-ray generation is buffered by battery packs which enable the generator to<br />

be powered by standard single phase AC power.<br />

The theory in this section is most easily understood while using the following<br />

schematics. Use the correct interconnect diagrams and circuit board schematics for your<br />

system.<br />

REFERENCE SCHEMATICS<br />

Analog Interface PCB<br />

Schematic # 00-876738 - All S/N<br />

Mainframe Motherboard<br />

Schematic # 00-875539 - System S/N 69-0001 thru 69-2000<br />

Schematic # 00-878396 - System S/N 69-2001 and up<br />

X-ray Regulator PCB<br />

Schematic # 00-877458 - All S/N<br />

Generator Driver PCB<br />

Schematic # 00-877461 - All S/N<br />

C-Arm Interconnect Diagram<br />

Schematic # 00-875500 - System S/N 69-0001 thru 69-1000<br />

Schematic # 00-877972 - System S/N 69-1001 thru 69-2000<br />

Schematic # 00-878376 - System S/N 69-2001 and up