Workstation Software 3<br />

DIAGNOSTICS<br />

Diagnostic tests are available for use in evaluating the operation of the Workstation<br />

electronics. The tests selectable from the System Diagnostics Menu are described in<br />

the following pages.<br />

1. Press the SETUP OPTIONS key on the Workstation. The Setup Options screen<br />

will appear.<br />

2. Insert the <strong>OEC</strong> Boot and Diagnostic disk into the disk drive.<br />

3. Select Access Level 2 and press the ENTER key.<br />

4. Select Run System Diagnostics and press ENTER.<br />

This screen appears when "Run System Diagnostics" is selected from the Service<br />

Options Menu.<br />

The tests available from this menu are described below:<br />

RUN SYSTEM DIAGNOSTICS<br />

Front Panel Test<br />

System Memory Tests<br />

Image Processor Tests<br />

Toggle Motion Artifact<br />

Use the cursor arrow keys to select<br />

option, then press ENTER.<br />

Press the ESC key to exit<br />

and return to previous screen.