Relation between Chart Speed and Intervals of Periodic Printout:<br />

Pen Model Dot Model<br />

IM 4H3B1-01E<br />

Chart Speed Printing Interval of Periodic Printout<br />

5 to 9 mm/h<br />

10 to 18 mm/h<br />

20 to 36 mm/h<br />

40 to 72 mm/h<br />

75 to 135 mm/h<br />

150 to 180 mm/h<br />

200 to 320 mm/h<br />

360 to 1500 mm/h<br />

more than 1500 mm/h<br />

no printout<br />

every 8 hours<br />

every 4 hours<br />

every 2 hours<br />

every hour<br />

every 30 minutes<br />

every 20 minutes<br />

every 10 minutes<br />

no printout<br />

DISPLAY<br />

Display Method: VFD (5 × 7 dot matrix, 20 characters).<br />

Digital Display: AUTO Channel No. or tag, type of alarm, measured values, unit (front 3<br />

digits), for each channel alternately<br />

MAN Channel No. or tag, type of alarm, measured values, unit (front 3<br />

digits), for one specific channel<br />

CLOCK Year / month / date and hour / min / sec will be displayed<br />

VIEW Status for IC Memory Card and TLOG function will be displayed<br />

Bargraph Display: Measured data Left-centered (%) bargraph and center-zero bargraph<br />

(selectable for each channel).<br />

Alarm Setting level of alarm, in case of alarm: flashing display.<br />

Other Displays: RCD: recording in progress<br />

POC: Pen Offset Compensation (pen model only)<br />

SET: SET Mode<br />

AUTO: Auto display is used<br />

MAN: <strong>Manual</strong> display is used<br />

CLK: Clock display is used<br />

ALM: shared alarm (not corresponding to any channel)<br />

CHT: Chart end indicator (optional)<br />

BAT: low battery<br />

Chart Speed Printing Interval of Periodic Printout<br />

6 dot 12 dot 18 dot 24 dot<br />

1 to 9 mm/h no printout no printout no printout no printout<br />

10 to 19 mm/h every 8 hrs. every 12 hrs. every 12 hrs. every 24 hrs.<br />

20 to 39 mm/h every 4 hrs. every 8 hrs. every 8 hrs. every 12 hrs.<br />

40 to 50 mm/h every 2 hrs. every 4 hrs. every 4 hrs. every 8 hrs.<br />

51 to 79 mm/h every 2 hrs. every 4 hrs. no printout no printout<br />

80 to 100 mm/h every hour every 2 hrs. no printout no printout<br />

101 to 1500 mm/h no printout no printout no printout no printout<br />

ALARM<br />

Number of Levels: Up to four levels for every channel (High, Low, Delta high, Delta low, High rate-ofchange,<br />

Low rate-of-change selectable)<br />

Interval time of rate-of-change alarm is the measuring interval multiplied by 1 to 15<br />

(selectable, and common for both rate-of-change alarms).<br />

Display: Set value: Is indicated as a segment on the bargraph.<br />

In case of alarm: This segment will start flashing on the bargraph and the digital<br />

display will show the type of alarm. The alarm indicator in<br />

different display will show alarm.<br />

Hysteresis: Approx. 0.5% of recording span (only High, Low) and 0%, selectable (common for all<br />

channels and all levels).<br />

10 - 5<br />

10. GENERAL SPECIFICATIONS