9.2.4 How to Select the Input Filter (Pen Model)<br />

Noise in the measured signal can be suppressed in the final measurement by using an<br />

input filter. This is sometimes referred to as signal damping. The purpose of this setting<br />

is to select for each channel whether or not to use a filter, and if so, to select a time<br />

constant of damping.<br />

A filter works as follows:<br />

meas.<br />

values<br />

Filter OFF<br />

Filter<br />

ON<br />

2, 5 or 10 sec.<br />

If a filter is used, the response time can be selected from 2, 5 or 10 seconds.<br />

You can choose from three filters:<br />

• 2 seconds<br />

• 5 seconds<br />

• 10 seconds.<br />

The initial value is ‘OFF’.<br />

SETTING PROCEDURE:<br />

Enter the SET UP Mode by turning ‘ON’ the power while pressing and holding the<br />

ENT-key until the SET UP display appears.<br />

Use the UP/DOWN-keys to select the display ‘SETUP=FILTR’. Press the ENT-key.<br />

Select the channel No. for which you want to specify a filter. Then press the ENT-key.<br />

Select the type of filter you want to use. You can choose from no filter (OFF), 2<br />

seconds, 5 seconds or 10 seconds. Press the ENT-key. The display ‘*FILTR SET*’ will<br />

appear.<br />

The filter setting has been completed, but has not been stored yet. Press the ENT-key to<br />

continue the setting for other channels or the ESC-key to adjust other settings in the<br />

SET UP Mode.<br />

Before leaving the SET UP Mode, you have to store your new settings. Press the ESCkey<br />

to return to the ‘SETUP=FILTR’ display and then select the ‘SETUP=END’<br />

display.<br />

Press the ENT-key. Select ‘END=STORE’ to keep your new settings or<br />

‘END=ABORT’ and press the ENT-key. After a few seconds, the Operation Mode will<br />

appear.<br />

9 - 12 IM 4H3B1-01E<br />

63.2%