Chapter 3 INSTALLATION<br />
3.1 Installation Site/Mounting<br />
IM 4H3B1-01E<br />
3.1 Installation Site/Mounting<br />
This chapter describes the installation of the µR1800, like the installation site, the<br />
mounting and wiring. To install the µR1800 properly, it is necessary to read this<br />
chapter.<br />
3.1.1 Installation Site<br />
The µR1800 installation site should be chosen to meet the following conditions as close<br />
as possible:<br />
Installation site should be panel-mounted only. The µR1800 is especially designed for<br />
panel mounting.<br />
The µR1800 should be placed in a well ventilated area.<br />
The µR1800 will be adversely affected if the unit is exposed to direct sunlight or<br />
installed near a heater. Choose a location near room temperature (23°C) with minimal<br />
temperature fluctuations.<br />
To expose the µR1800 to soot, steam, moisture, dust, corrosive gases etc. will adversely<br />
affect it.<br />
To use the µR1800 within strong electro-magnetic fields may cause malfunction. Please<br />
avoid installing near electro-magnetic objects.<br />
To install the µR1800 in a location susceptible to mechanical vibrations will adversely<br />
affect the mechanical parts and the quality of recording. Please choose an installation<br />
site characterized by minimal mechanical vibrations.<br />
This µR1800 must be installed horizontally (however, mounting may be inclined up to<br />
30° backwards from vertical).<br />
To use the µR1800 within domestic establishments and within establishments directly<br />
connected to a low voltage power supply network which supplies buildings used for<br />
domestic purposes may cause malfunction of other equipments. Please avoid using in<br />
domestic environment.<br />
3 - 1<br />
3. INSTALLATION