7.7 How to Set the Remote RJC (For /N5 Model)<br />

When measuring a temperature by thermocouple (TC) and the object of measurement is<br />

located at a big distance, it is possible to measure the temperature without using a lot of<br />

expensive TC wires by establishing a relay terminal near the measurement object.<br />

The relay terminal and object of measurement are connected by TC. The relay terminal<br />

and input terminal of the main unit are connected by a compensating copper wire.<br />

One other input terminal of the main unit is connected with the relay terminal by TC<br />

and by couducting standard setting compensation of the to be measured temperature,<br />

which is done by measuring the temperature of the relay terminal, the temperature of the<br />

measurement object can be calculated. (Refer to the figure below.)<br />

main unit relay terminal (to be prepared by customer)<br />

TC*<br />

A<br />

reference channel<br />

measurement channels<br />

compensating<br />

B TC*<br />

copper wire<br />

B<br />

compensating<br />

B TC*<br />

coppwer wire<br />

B<br />

input terminal<br />

* Please use the same type of TC<br />

NOTES The input mode of the reference channel should be same as the used TC.<br />

The range of each measuring channel will become same as the range of the reference<br />

channel.<br />

The limits of the span of the measuring channel are the same as for the usual TC input<br />

(refer to page 7-4).<br />

Scaling is not possible.<br />

In case you change the range or input mode of the reference channel, the input mode of<br />

the measurement channel being TC will not change, while the span becomes the upper/<br />

lower value of the newly set range.<br />