5.3.9 How to Cope with FAIL/Chart End<br />

When equipped with this option and the chart reaches its end or a CPU-error occurs, the<br />

output relay will be activated. There is one output relay for Chart end detection, and one<br />

for FAIL detection. For the wiring, see 3.2.5.<br />

CHART END:<br />

When there are approx. 2cm left till chart end, the ‘CHT’ indicator will be illuminated<br />

in the display and recording will stop. To change the chart, follow the steps as explained<br />

in 5.2.1.<br />

At the same time the output relay will be activated. This output relay is of the<br />

energizing type and this cannot be changed. See also the explanation mentioned below.<br />

FAIL:<br />

When a CPU-error occurs, the output relay will be activated. This relay is of the deenergizing<br />

type and this cannot be changed. Therefore this relay will also be activated<br />

when the power is off. The activating of this relay will not be shown on the display.<br />

See also the explanation mentioned below.<br />

When a CPU-error occurs, contact the nearest Sales & Service Office. Addresses may<br />

be found on the back cover of this manual.<br />

Explanation:<br />

Chart end output relay (energize type):<br />

NO C NC<br />

Normal operation<br />

NO C NC<br />

Chart end occurs<br />

FAIL output relay (de-energize type):<br />

NO C NC<br />

Normal operation<br />

NO C NC<br />

Chart end occurs<br />

(FAIL)<br />

NO C NC<br />

Power OFF<br />

NO C NC<br />

Power OFF<br />

5 - 16 IM 4H3B1-01E