11.5 Calibration<br />
IM 4H3B1-01E<br />
It is recommended to perform a calibration once a year to assure the measuring<br />
accuracy.<br />
Required instruments:<br />
DC Voltage Standard: YOKOGAWA Model 2552 or equivalent.<br />
Decade Resistance Box: YOKOGAWA Model 2793-01 or equivalent.<br />
11.5 Calibration<br />
Calibration procedure:<br />
1 Connect the calibration instrument to the input terminals of the recorder as shown in<br />
figures 11.3, 4 , 5 or 6 and allow the recorder to warm up for at least 30 minutes.<br />
Allow the instrument to warm up according to its specifications.<br />
2 Check that the ambient temperature and humidity are within Normal Operating<br />
Conditions (see chapter 10 ‘General Specifications’).<br />
3 Apply an input corresponding to 0, 50 and 100 percentage of the entered setting range,<br />
and calculate errors from the readings on the display.<br />
4 If the error is at any point outside the accuracy limits (for accuracies, see chapter 10<br />
‘General Specifications’), contact your nearest Sales & Service Office. Addresses may<br />
be found on the back cover of this manual.<br />
NOTE For a TC input, it is necessary to measure the temperature of the input terminals, and<br />
add a voltage corresponding to the reference temperature.<br />
DC Voltage Measurement<br />
The figure blow shows the pen model, but connection should be done in the same way<br />
for the dot model.<br />
Power Supply terminal<br />
(for others than /P1 model)<br />
1<br />
2<br />
3<br />
4<br />
– +<br />
L<br />
N<br />
Input terminals<br />
(DC V & TC inputs)<br />
Figure 11.3 DC Voltage Measurement<br />
DC Voltage Standard<br />
+<br />
G<br />
–<br />
11 - 5<br />
11. MAINTENANCE