IM 4H3B1-01E<br />
EXPLANATION OF SQUARE ROOT:<br />
The µR1800 uses the following square rooting-method:<br />
Let us define the items as follows:<br />
V min = minimum value of recording span (LEFT)<br />
V max = maximum value of recording span (RIGHT)<br />
F min = minimum value of scale (SCL LEFT)<br />
F max = maximum value of scale (SCL RIGHT)<br />
V x = input voltage<br />
F x = scaling value<br />
Fmax<br />
Fmin<br />
F<br />
a<br />
Vmin<br />
b<br />
(Fmax – Fmin ) x 10%<br />
b'<br />
(V max – V min ) x 1%<br />
c<br />
Vmax<br />
7.1 How to Set Input Range and Recording Span<br />
The relationship between V x (input voltage) and F x (scaling value) is as shown in the<br />
graph below (the graph configuration is approximate).<br />
Between b and c in the graph, the following relation exists between F x and V x:<br />
F = (F – F )<br />
V x – V min<br />
x max min<br />
+ F<br />
V max – V min<br />
min<br />
And between a and b, the relation is:<br />
10 (F<br />
F = max – F min)<br />
x (V x – V min ) + Fmin<br />
V max – V min<br />