27.06.2013 Views

Modeles 4370 UR1800 Recorder Instruction Manual

Modeles 4370 UR1800 Recorder Instruction Manual

Modeles 4370 UR1800 Recorder Instruction Manual

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

IM 4H3B1-01E<br />

The input value and measured value<br />

are very different.<br />

(e.g. pens are staying at 0% or 100%<br />

of the chart, big fluctuations.)<br />

Does the input satisfy the<br />

specifications? (signal source<br />

resistance, etc.)<br />

YES<br />

Are the span and range<br />

settings correct?<br />

YES<br />

Is the input free of noise?<br />

YES<br />

Is the<br />

environmental temperature<br />

constant? (in case of temperature<br />

measurement)<br />

YES<br />

Is the input wiring correct?<br />

YES<br />

Is the recorder<br />

free of parallel wiring with<br />

other equipment?<br />

YES<br />

RJC INT/EXT setting is<br />

correct?<br />

(TC input)<br />

YES<br />

Is position of pens or<br />

printer carriage correct?<br />

YES<br />

NO<br />

NO<br />

NO<br />

NO<br />

NO<br />

NO<br />

NO<br />

Contact your nearest Sales & Service center<br />

12.2 Troubleshooting Flow Chart<br />

Meet the specifications<br />

Adjust the span and range<br />

Ground the recorder properly.<br />

Ground the measuring object properly.<br />

Insulate TC and measuring object.<br />

Use shielded wire for input.<br />

Correct A/D integration time (see 9.2.1).<br />

Use an input filter (see 9.2.4).<br />

Separate noise generating wire.<br />

Use an input terminal cover.<br />

Protect input terminal from wind or fan.<br />

Stabilize room temperature around the<br />

input terminal.<br />

Wire correctly (see 3.2.3).<br />

Fix the terminal block again.<br />

Fasten terminal screws.<br />

Insulate RTD from ground.<br />

If broken, replace TC.<br />

Stop burnout function of other equipment.<br />

Ground the recorder and other equipment<br />

at the same grounding place.<br />

Disconnect parallel wiring.<br />

(e.g. using double element TC)<br />

Set correctly (see 9.2.3)<br />

NO Adjust pen or printer carriage<br />

position (see 11.3 or 11.4)<br />

Adjust the position of the<br />

printer carriage (see 11.3)<br />

12 - 3<br />

12. TROUBLESHOOTING

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!