TABLE-1 : EFFECTIVE FOCAL SPOT SIZE FOR<br />

A MAGNIFICATION OF 4/3<br />

Smallest group<br />

resolved<br />

1<br />

2<br />

3<br />

4<br />

5<br />

6<br />

7<br />

8<br />

9<br />

10<br />

11<br />

12<br />

lp/mm<br />

0.84<br />

1.00<br />

1.19<br />

1.14<br />

1.68<br />

2.00<br />

2.38<br />

2.83<br />

3.36<br />

4.00<br />

4.76<br />

5.66<br />

Effective focal spot size<br />

(mm)<br />

4.3<br />

3.7<br />

3.1<br />

2.6<br />

2.2<br />

1.8<br />

1.5<br />

1.3<br />

1.1<br />

0.9<br />

0.8<br />

0.7<br />

TABLE 2 : HVT AS A FUNCTION OF FILTRATION AND TUBE<br />

POTENTIAL (SINGLE PHASE GENERATORS)<br />

Total<br />

filtration<br />

(mm Al)<br />

0.5<br />

1.0<br />

1.5<br />

2.0<br />

2.5<br />

3.0<br />

3.5<br />

Peak potential (kV)<br />

30 40 50 60 70 80 90 100 110 120<br />

Half Value Thickness (mm Al)<br />

0.36 0.47 0.58 0.67 0.76 0.84 0.92 1.00 1.08 1.16<br />

0.55 0.78 0.95 1.08 1.21 1.33 1.46 1.58 1.70 1.82<br />

0.78 1.04 1.25 1.42 1.59 1.75 1.90 2.08 2.25 2.42<br />

0.92 1.22 1.49 1.70 1.90 2.10 2.28 2.48 2.70 2.90<br />

1.02 1.38 1.69 1.95 2.16 2.37 2.58 2.82 3.06 3.30<br />

1.49 1.87 2.16 2.40 2.62 2.86 3.12 3.38 3.65<br />

1.58 2.00 2.34 2.60 2.86 3.12 3.40 3.68 3.95<br />

245