24.11.2014 Views

Volume 2 - Atomic Energy Regulatory Board

Volume 2 - Atomic Energy Regulatory Board

Volume 2 - Atomic Energy Regulatory Board

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Result :<br />

Coefficient of linearity for timer of the unit is ___________<br />

Tolerance : Coefficient of linearity < 0.1<br />

9.2 Output Consistence:<br />

Operating mAs X, Instrument reading (mR) Avg. mR/mAs<br />

potential ξ M<br />

1 2 3 4 5<br />

Tolerance : Coefficient of variation < 0.05<br />

9.3 Radiation Leakage Levels from X-ray Tube Housing at 1 metre from the<br />

Focus<br />

Operating potential : ________ kV, mAs : _______ mA x ______seconds<br />

Radiation leakage level (mR/h or mR)<br />

Cathode Anode Stand Front Top<br />

Result : Maximum radiation leakage at 1 metre from the focus for workload<br />

of 180 mA<br />

minutes in one hour is ___________ mR.<br />

Recommended upper limit :<br />

Leakage radiation level at 1 metre from the focus<br />

≤ 115 mR in one hour.<br />

9.4 Radiation Leakage Levels from the Diaphragm/Collimator at 1 metre from<br />

the Focus<br />

Operating potential : ________ kV, mAs : ________ mA x ________seconds<br />

Radiation leakage level (mR/h or mR)<br />

Cathode Anode Stand Front Top<br />

Result : Maximum radiation leakage through diaphragm/collimator at 1 metre<br />

from the focus for work-load of 180 mA minutes in one hour is<br />

____________ mR.<br />

252