Guidelines for care & Use of Dry Solvent Stills [Example]
Guidelines for care & Use of Dry Solvent Stills [Example]
Guidelines for care & Use of Dry Solvent Stills [Example]
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>Use</strong> <strong>of</strong> Reflux Condensers<br />
Both undergraduate organic laboratories and research laboratories frequently require the heating <strong>of</strong><br />
liquid mixtures to facilitate chemical reactions. These mixtures are composed <strong>of</strong> reactants dispersed in<br />
organic solvents, usually in round-bottomed flasks capped by upright, jacketed water-filled glass tubes<br />
refereed to as reflux condensers.<br />
Common, generally nontoxic solvents can be refluxed on open laboratory benches. Ideally,<br />
however, all reflux condensers should be set up in a hood, in case too much heat is applied and vapors rise<br />
out <strong>of</strong> the condenser. Since boiling solvents do not increase in temperature, increasing the heat applied will<br />
only cause loss <strong>of</strong> solvent vapor, not an increase in reaction rate. A condensation level <strong>of</strong> 1 or 2 inches<br />
above the bottom <strong>of</strong> the condenser tube is all that is necessary.<br />
Tap water is cold enough <strong>for</strong> most refluxes. However, very low boiling solvents like acetone or<br />
ethyl ether may require ice-cooled water supplied with submersible lab pumps. Constant temperature waterbaths<br />
can also be used.<br />
The condensers should be placed vertically onto the round bottom heating flask and attached<br />
firmly to ring stands with metal clamps. The water inlet should be through the lower end <strong>of</strong> the condenser.<br />
Water hoses should be clamped onto water inlets/outlets on the condensers with Ace Glass wire hose<br />
clamps (located in the stockroom <strong>for</strong> Departmental use) or thick copper wire, twisted on firmly with a pair<br />
<strong>of</strong> pliers (Thin wire could eventually slice through the tubing). Hose clamps are not a good idea. Surging<br />
pressure in Salem Hall water lines could cause water hoses to slip <strong>of</strong>f if they are held onto a serrated<br />
condenser hose adapter fittings with flat hose clamps. The only disadvantage <strong>of</strong> using wires is that you<br />
must cut them <strong>of</strong>f periodically when they become corroded and replace them. <strong>Use</strong> <strong>of</strong> Tygon tubing, instead<br />
<strong>of</strong> rubber, will allow you to actually see the water flowing through the hose. You can then run water<br />
through with a relatively slow speed, which is all you need in water condenser coils. Also, this means there<br />
will be less chance <strong>of</strong> a spill when inevitable water pressure surges occur in Salem Hall at night. The water<br />
pressure never falls to zero, so you basically can’t have the water flow on too low. Once you have<br />
connected the water hose clamps, gently but firmly pull on the lines to make sure they are on tight.<br />
I am currently looking <strong>for</strong> inexpensive in-line water flow indicators (will purchase those without<br />
molded plastic halves which could pry apart under water pressure surges) which you can attach anywhere in<br />
the water line to more readily judge the flow <strong>of</strong> water through the line.<br />
You may also purchase Water-Flow Monitor sensor instruments with solenoid valve automatic<br />
water shut-<strong>of</strong>f mechanisms <strong>for</strong> about $600 from I2R Company.<br />
"Bumping" <strong>of</strong> solvents can be avoided by adding boiling stones.<br />
Magnetic stir bars can also be added to stir the boiling solution, allowing more even boiling.<br />
Safe <strong>Use</strong> <strong>of</strong> Laboratory Centrifuges<br />
Lab centrifuges in Salem Hall are <strong>of</strong> the low-speed type, and do not need special handling<br />
requirements. They are supported on suction-cup feet and will not ordinarily wobble to the point <strong>of</strong> falling<br />
<strong>of</strong>f bench tops. Balance them by placing test tubes half filled with water immediately opposite the test tube<br />
being centrifuged.<br />
Occasionally check the power cord and keep extraneous lab equipment from cluttering bench tops<br />
near the centrifuge.<br />
136