OPERATING INFORMATION FOR FHM 410 FUME HOOD MONITORS<br />

LOCATED ON RESEARCH LAB HOODS<br />

- Green lights indicate hood is safe to use.<br />

- Red lights indicate danger -- hood is not safe. Call maintenance at phone<br />

#4255, ask <strong>for</strong> Donnie Adams, Building Zone Supervisor.<br />

- No lights indicate loss <strong>of</strong> power. Call maintenance at phone #4255.<br />

EMERGENCY EXHAUST<br />

LIGHT (RED)<br />

Indicates the emergency exhaust<br />

mode is activated. See Emergency<br />

Button below <strong>for</strong> an explanation <strong>of</strong><br />

the Emergency Exhaust Mode.<br />

EMERGENCY BUTTON (RED)<br />

USE WHEN A HAZARDOUS<br />

MATERIAL SPILL OCCURS IN<br />

HOOD OR NEARBY LAB<br />

AREA.<br />

Push this button to activate the<br />

emergency exhaust mode. An<br />

alarm will sound. In this mode<br />

exhaust air is at its maximum flow<br />

to rapidly evacuate the hood. (The<br />

red light in the switch indicates the<br />

switch is turned on). Push the<br />

button again to turn <strong>of</strong>f the<br />

emergency exhaust mode.<br />

FUME HOOD MONITOR<br />

SYSTEM STATUS<br />

STANDBY OPERATION<br />

-NORMAL-<br />

STANDARD OPERATION<br />

EMERGENCY EXHAUST<br />

CAUTION - FLOW ALARM<br />

EMERGENCY MUTE<br />

Automatic constant Face Velocity <strong>for</strong> Safety<br />

PHOENIX CONTROLS CORPORATION<br />

FHM 410 Model<br />

NORMAL LIGHTS (GREEN)<br />

Indicates normal operation <strong>of</strong><br />

fume hood. One <strong>of</strong> these lights<br />

must be on <strong>for</strong> the hood to be safe<br />

to use.<br />

Standard Operation: system<br />

operating at standard face<br />

velocity<br />

Standby Operation: system<br />

operating at a lower face velocity<br />

(active only with Zone Presence<br />

Sensor, which are not presently<br />

installed).<br />

CAUTION FLOW ALARM<br />

LIGHT (RED)<br />

Indicates unsafe airflow<br />

condition. An alarm will sound.<br />

Immediately close sash and call<br />

maintenance at phone #4255.<br />

DO NOT USE THE HOOD.<br />

MUTE SWITCH (GRAY)<br />

Audible alarms are energized on<br />

any <strong>of</strong> the following conditions:<br />

- CAUTION - FLOW ALARM or<br />

- EMERGENCY EXHAUST<br />

Push the button to silence alarm.<br />

Alarm status lights will remain<br />

lit. The mute mode is reset<br />

automatically after alarm<br />

conditions clear.<br />