12.07.2015 Views

User's Manual - Cornell Lab of Ornithology - Cornell University

User's Manual - Cornell Lab of Ornithology - Cornell University

User's Manual - Cornell Lab of Ornithology - Cornell University

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Chapter 4: Amplitude CalibrationSelectingcalibrationparametersPressureSelect two calibration parameters by clicking two radio buttons next to theterms Pressure, Intensity and Characteristic Impedance. The values shown forthe two parameters that you select can be edited. The value <strong>of</strong> the thirdparameter is calculated by Canary from the other two, according to Equation4.1 or 4.2.The pop-up menu next to the value shown for Pressure indicates whetherpressure is expressed in absolute pressure units (pascals, abbreviated Pa) or indecibels (dB). If you choose dB, the value in the Pressure field is taken as dBrelative to the pressure reference value specified in the Pressure dB re field inthe lower left part <strong>of</strong> the dialog box (see below). You should select the unitsbefore entering a value into the Pressure field.A second pop-up menu lets you specify whether the value in the Pressurefield will be the RMS, Peak, or Ceiling value for pressure. Figure 4.2 illustratesthe relationship between Ceiling, Peak, and RMS values. If you choose RMS,the value is taken as the RMS pressure value <strong>of</strong> the entire signal. If you choosePeak, the value is the pressure value <strong>of</strong> the individual sample with the largestabsolute value. If you choose Ceiling, the value is the pressure represented bya sample in which all bits are equal to 1. 1 You should select RMS, Peak orCeiling before entering a value into the Pressure field.peak = 1.58 Paceiling = 2.0 PaRMS = 0.78 PaFigure 4.2. Relationship between ceiling, peak, and RMS values <strong>of</strong> a signal.Canary lets you assign an absolute amplitude value to any <strong>of</strong> these threevalues. This signal has been calibrated to set the ceiling equal to 2.0 Pa (=100dB re 20 µPa).If you selected Pressure as one <strong>of</strong> the two parameters to be set (by clicking oneither <strong>of</strong> its radio buttons), enter the RMS, peak, or ceiling pressure <strong>of</strong> theentire signal. If you are using a pure-tone calibration signal <strong>of</strong> knownamplitude, as discussed above, you will probably want to specify the RMS orpeak pressure.If you did not select Pressure as one <strong>of</strong> the two values to be set, the pressureshown will be the pressure that Canary has calculated from the current valuesfor Intensity and Impedance.1 The ceiling value thus corresponds to 128 for 8-bit samples, or 32768 for 16-bitsamples.Canary 1.2 User’s <strong>Manual</strong> 73

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!