7.<br />

Kassér kanyler og sprøjter ifølge etableret protokol.<br />

B. Skylningsskema<br />

Systemets integritet skal bekræftes, inden porten skylles igennem (se Bestemmelse<br />

af systemintegritet).<br />

Ved venøse systemer bevares systemets åbenhed ved at det skylles med en<br />

heparinopløsning mindst en gang om måneden når systemet ikke er i brug.<br />

NB: <strong>ProPort</strong>-systemets selvforseglende membran tillader op til 2000<br />

punkturer med en 0,7 mm PORT-A-CATH®-nål. Den selvforseglende membran<br />

på en <strong>ProPort</strong> Low profile port tillader op til 1500 punkturer med en 0,7 mm<br />

PORT-A-CATH®-nål.<br />

C. Procedure for blodprøver<br />

Systemets integritet skal bekræftes, inden en blodprøveprocedure udføres (se<br />

Bestemmelse af systemintegritet).<br />

NB: Det anbefales at anvende en 10 ml eller større sprøjte til injektion eller<br />

infusion af væsker igennem systemet.<br />

1.<br />

2.<br />

3.<br />

4.<br />

5.<br />

6.<br />

Skyl systemet igennem med en ikke-udkernende PORT-A-CATH® adgangskanyle<br />

med 10 ml fysiologisk saltvand for at bekræfte, at systemet ikke er<br />

okkluderet.<br />

Tag langsomt mindst 5 ml blod og kassér det.<br />

Tag den ønskede mængde blod.<br />

Injicér øjeblikkeligt 20 ml fysiologisk saltvand.<br />

Dan en blokering med heparinopløsningen ved at instillere 5 ml heparinopløsning<br />

(10 til 100 IU/ml). Oprethold overtrykket ved at trække sprøjten<br />

tilbage samtidig med, at de sidste 0,5 ml heparinopløsning injiceres.<br />

Kassér kanyler og sprøjter ifølge etableret protokol.<br />

D. Indstilling af brug af systemet<br />

Hvis det bestemmes, at systemet ikke længere er nødvendigt for behandling,<br />

bør lægen overveje eksplantation af systemet. Hvis systemet skal blive siddende<br />

anbefales periodiske røntgenoptagelser med patienten i opret stilling og med<br />

armene ved siden. Dette muliggør registrering af problemer med systemet, som<br />

f.eks. sammenklemning af katetret mellem kravebenet og første ribben, hvilket<br />

kan resultere i kateterfragmentering og efterfølgende embolisering. 1,4,5,8<br />

E. Litteraturhenvisninger<br />

1.<br />

2.<br />

3.<br />

4.<br />

5.<br />

6.<br />

7.<br />

8.<br />

9.<br />

10.<br />

Klotz HP, Schopke W, Kohler A, Pestalozzi B, Largiader F. Catheter Fracture: A Rare<br />

Complication of Totally Implantable Subclavian Venous Access Devices. J Surgical<br />

Oncol 1996; 62:222.<br />

Vadlamani P, Dawn B, Perry MC. Catheter Fracture and Embolization from Totally<br />

Implanted Venous Access Ports. J Vascular Dis 1998; 12:1013.<br />

Monsuez JJ, Douard MC, Martin-Bouyer Y. Catheter Fragments Embolization. J<br />

Vascular Dis 1997; 2:117.<br />

Ramsden WH, Cohen AT, Blanshard KS. Case Report: Central Venous Catheter Fracture<br />

Due to Compression Between the Clavicle and First Rib. Clinical Radiology 1995; 50:59.<br />

Coles CE, Whitear WP, Le Vay JH. Spontaneous Fracture and Embolization of a Central<br />

Venous Catheter: Prevention and Early Detection. Clinical Oncology 1998; 10:412.<br />

Kossoff EH, Poirier MP. Peripherally inserted central venous catheter fracture and<br />

embolization to the lung. Pediatric Emer Care 1998; 6:<strong>40</strong>3.<br />

Lafreniere R. Indwelling Subclavian Catheters and a Visit With the “Pinched-Off Sign”.<br />

J Surgical Oncol 1991; 47:261.<br />

Hinke DH, Zandt-Stastny DA, Goodman LR, Quebbeman EJ, Krzywda EA, Andris<br />

DA. Pinch-off Syndrome: A Complication of Implantable Subclavian Venous Access<br />

Devices. Radiology 1990; 177:353<br />

Franey T, Demarco LC, Geiss AC, Ward RJ. Catheter fracture and embolization in a<br />

totally implanted venous access catheter. J Parent Ent Nutr 1988; 12:528<br />

Moureau NL, McKinnon BT, Douglas CM. Multidisciplinary management of thrombotic<br />

catheter occlusions. J Vasc Acces Dev 1999; summer:22<br />

NB: En udgivelsesdato for denne brugsanvisning er inkluderet for lægens<br />

information. Lægen bør, hvis der er gået mere end et år mellem udgivelsesdatoen<br />

og anvendelsen af produktet, kontakte Deltec, Inc. for at forhøre, om<br />

en nyere revideret udgave af brugsanvisningen er tilgængelig.<br />

Udgivelsesdato: 2005-06<br />

De beskrevne produkter er beskyttet af ét eller flere af følgende U.S.-patentnumre:<br />

4,723,948, 4,743,231, 4,880,414, 4,963,133, 5,387,192, 5,558,641 og 5,613,945. Andet/andre<br />

patent(er) anmeldt. Udenlandsk(e) patent(er) anmeldt.<br />

<strong>ProPort</strong>, SlideLock, CATH-SHIELD, PORT-A-CATH og Low Profile er varemærker, der<br />

tilhører familiekoncernen <strong>Smiths</strong> <strong>Medical</strong>. Symbolet angiver, at produktet er registreret<br />

i U.S. Patent and Trademark Office (amerikanske bureau ® for patenter og varemærker) og i<br />

visse andre lande.<br />

© 2005 <strong>Smiths</strong> <strong>Medical</strong> familiekoncernen. Alle rettigheder forbeholdt.<br />