Y scale unit<br />
Selects the measurement units. It is possible to select Voltage (dBV, dBu, dBRel) or<br />
Pressure (dBSPL) or Impedance (Ohm).<br />
smoothing<br />
Activates a frequency smoothing of the active curve.This smoothing effect will allow<br />
a better appreciation of the general features of the response curve. The smoothing<br />
algorithm that is employed averages all values within a fraction-of-octave band<br />
surrounding each analysis frequency.<br />
10.2.3 MLS & LOG CHIRP SETTINGS DIALOG<br />
Figure 10.2<br />
sampling<br />
Selects the measurement sampling frequency. When LOG CHIRP is selected as<br />
stimuli, 48kHz only is available.<br />
size<br />
Selects the size of the MLS & LOG CHIRP sequence.<br />
window<br />
Selects the appropriate kind of window for analyzing time data. It is possible to select<br />
between a rectangular, Hanning or Blackman window; the last two can be full or half<br />
sized. NOTE: These windows are applied to the time portion to be transformed with<br />
FFT. If the start point is near the impulse, full windows will null the most important<br />
part of the time response (due to their rise time). To evaluate the effects of a data<br />
window refer to Chapter 9, and FFT measurements in general.<br />
stimuli<br />
Selects the kind of stimulus, either MLS and LOG CHIRP used for the measurement.<br />
averages<br />
Controls the averaging mode of operation. The measurement will be repeated and<br />
averaged the number of times set, therefore obtaining a better signal-to-noise ratio<br />
at the expense of reduced measurement speed. Continuous performs the number<br />
of averages in the shortest time without waiting. <strong>Manual</strong> waits the user to press any<br />
key between each measure; it is useful, for example, in averaging different<br />
microphone positions.<br />
impedance<br />
Set how Impedance is calculated. When taking impedance measurements refer<br />
either to the internal impedance mode or to the ISense feature of the CLIOQC<br />
Chapter 10 - MLS 111