RTU [s]. Reverberation time evaluated from a user defined dynamic range; refer to<br />

15.3 acoustical parameters settings. See also below the correlation coefficient R<br />

associated with RTUser.<br />

R(RT). Each reverberation time estimation (RT20, RT30 and RTU) has associated a<br />

negative number which is the correlation coefficient R showing how closely the<br />

corresponding decay curve fits a straight line. A value of -1 gives a perfect linear fit.<br />

When the correlation coefficient is smaller than -0.95 the RT value should be<br />

viewed with suspect as the decay curve may not be sufficiently linear; direct<br />

inspection of the decay curve with markers should be carried out.<br />

182 Chapter 15 - Acoustical Parameters