CLIOwin 7 PCI User's Manual - Audiomatica
CLIOwin 7 PCI User's Manual - Audiomatica
CLIOwin 7 PCI User's Manual - Audiomatica
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
11.3 A BRIEF DESCRIPTION ON SETTINGS EFFECTS<br />
11.3.1 STEPPED VS. NOT STEPPED<br />
Although measuring speed increases, use of a “not stepped” sweep can adversely affect<br />
measuring results in several circumstances. As an example that should make this clear,<br />
let’s see what happens while measuring the impedance of a woofer in Internal or<br />
Constant Current Mode. Please refer to Measuring Impedance for more information on<br />
this topic. In both conditions the loudspeaker is driven from a high impedance source<br />
and its damping is only mechanical. Fig.11.1 shows a 6" woofer, driven by a 100Ohm<br />
output impedance generator delivering a sinusoidal burst 200ms long, at its resonance<br />
frequency. When the excitation stops, the device continues to move, and therefore<br />
produce back electromotive force (EMF) voltage for more than 50 ms. Something very<br />
similar happens at start-up. In this situation, if CLIO is set in “not stepped” mode, it will<br />
acquire this EMF together with the actual results.<br />
0.200<br />
CLIO<br />
V<br />
0.120<br />
0.040<br />
-0.040<br />
-0.120<br />
-0.200<br />
0.00 50 100 150 200 250 300 350 400 ms 450 500<br />
Figure 11.1<br />
Fig.11.2 shows the same driver impedance taken both in stepped (black) and “not<br />
stepped” mode (red). The “not stepped” curve is simply wrong. Conclusion: use always<br />
stepped modeor the highest frequency resolution if the behavior of the device to be<br />
measured is unknown.<br />
25.0<br />
CLIO<br />
180.0<br />
Ohm<br />
Deg<br />
20.0 108.0<br />
15.0 36.0<br />
10.0 -36.0<br />
5.0 -108.0<br />
0.0 -180.0<br />
10 100 1k Hz<br />
10k 20k<br />
Figure 11.2<br />
136 Chapter 11 - Sinusoidal