memory or saved measurements. Ohm switches the system to convert the<br />

measurements in Ohm basing the conversion on the Impedance Mode Settings<br />

available in the Settings Dialog. Again, once the measurements have been taken the<br />

data is immediately converted.<br />

Smoothing<br />

Allows the user to select a Frequency smoothing of the active curve. The smoothing<br />

algorithm averages all the value within the selected fraction of octave band,<br />

surrounding each analysis Frequency. It is a non destructive post process that can<br />

be applied or removed at any moment after the measurement has been taken.<br />

11.2.3 SINUSOIDAL SETTINGS DIALOG<br />

This is undoubtedly the most important User Interface of the entire Menu. Here the single<br />

action of every control is considered. Theoretical and practical effects on the subsequent<br />

measurement are considered later.<br />

Sweep Settings<br />

Stepped Check Box<br />

Allows the user to choose between continuous or stepped Logarithmic Sweep.<br />

Continuous sweep is faster but, therefore, should only be used if it’s clear to the user<br />

how this affects the measurements.<br />

Resolution Drop Down<br />

Lets the user choose between five different Frequency resolutions. It affects the<br />

measurement execution time, whatever the other settings are.<br />

Freq Max Edit Box<br />

Lets the user define the highest Frequency of the Sweep. This is also the starting<br />

measurement Frequency. The highest value accepted is 22388 Hz. The value should<br />

also be one octave higher than the Minimum Frequency.<br />

Freq Min Edit Box<br />

Lets the user define the lowest Frequency of the Sweep. This is also the end<br />

measurement Frequency. The lowest value accepted is 10 Hz. The value should also<br />

be one octave lower than the Maximum Frequency.<br />

Chapter 11 - Sinusoidal 133