9.4 FFT SETTINGS DIALOG<br />

Fig. 9.2 The FFT settings dialog box<br />

FFT Size<br />

Selects the number of samples acquired and processed by each FFT. It is possible<br />

to choose a size between 512 and 131072 points.<br />

Sampling<br />

Selects the sampling frequency.<br />

Delay<br />

Permits the input of the desired processing delay (in ms) when in Internal Trigger<br />

mode. See 9.7 for details.<br />

Internal Trigger<br />

Enables the Internal Trigger mode. See 9.4 for details.<br />

Enable Frequency Calibration<br />

Enables frequency calibration to compensate for any hardware non linearity;<br />

frequency calibration, if enabled, takes place only when the generator is active.<br />

Hold Function<br />

Selects either Min or Max hold function. This functionality is activated by the relative<br />

toolbar button.<br />

Freq Axis<br />

Selects from linear or logarithmic frequency axis (valid only for FFT narrowband)<br />

Enable Equal Loudness Contour<br />

Enables the display of the normal equal loudness level curves as defined in the ISO<br />

226 standard. The curves are displayed only in FFT narrowband and RTA modes when<br />

dBSPL units are selected.<br />

Averaging<br />

Selects either linear or logarithmic averaging; see Averaging (9.6) for details.<br />

96 Chapter 9 - FFT