8.6 INTERACTION BETWEEN THE MULTI-METER AND FFT<br />
The Multi-meter uses the same capture and processing units as the FFT control panel.<br />
To perform a measurement it programs the FFT routines (changing FFT settings to<br />
match its needs) and then effectively starts an FFT measurement in background.<br />
The two panels can be opened and can work together but FFT always acts as the master<br />
while Multi-meter as the slave. In this situation the Multi-meter window's title is changed<br />
to 'Multi-Meter (FFT slave)' to reflect this new state; the Go and Stop buttons are<br />
disabled as you operate the slave panel from inside FFT, starting and stopping the<br />
reading with the FFT's Go and Stop buttons; the input channel follows the FFT one<br />
while the integration setting become meaningless; this is because the user has<br />
control over FFT averages which precisely define the measurement integration.<br />
Nevertheless, during slave operation, it is possible to select the displayed parameter and<br />
its unit. It is not possible to select LCR operation as the LCR meter uses different<br />
measurement capabilities.<br />
As soon as the FFT control panel is closed it releases Multi-meter from the slave state;<br />
then the Multi-meter is then ready to operate in a stand-alone capacity and is fully<br />
functional as described above.<br />
92 Chapter 8 - Multi-Meter