Table 28-3. Laboratory Quality Control Samples – Water/EPA Method 6010B<br />

Matrix Water<br />

Analytical<br />

Group<br />

Metals<br />

Method/SOP<br />

EPA 6010B<br />

Reference<br />

QC Sample<br />

Method<br />

Blank<br />

Frequency/<br />

Number 1<br />

1/batch<br />

Method/SOP<br />

QC Acceptance<br />

Limits<br />

No detects ≥ ½<br />

QL<br />

LCS 1/batch 80%-120%<br />

Postdigestion<br />

Spike<br />

Serial<br />

Dilution<br />

When dilution<br />

fails<br />

1/batch<br />

75%-125%<br />

≤10% <strong>for</strong><br />

analytes with<br />

concentration<br />

>50x QL<br />

MS/MSD 1/batch 80%-120%<br />

Method of<br />

Standard<br />

Addition<br />

When matrix<br />

interference is<br />

suspected<br />

Corrective<br />

Action<br />

Reanalyze with<br />

all associated<br />

samples<br />

Reprep and<br />

reanalyze the<br />

LCS and all<br />

samples<br />

Analyze<br />

samples by<br />

method of<br />

standard of<br />

addition<br />

Per<strong>for</strong>m post<br />

digestion spike<br />

If the result is<br />

indicative of<br />

matrix<br />

interference,<br />

discuss in the<br />

case narrative.<br />

Otherwise,<br />

check <strong>for</strong> a<br />

possible source<br />

of error, and<br />

extract/reanalyz<br />

e the sample.<br />

Person(s)<br />

Responsible<br />

<strong>for</strong> Corrective<br />

Action<br />

Analyst<br />

Data Quality<br />

Indicator<br />

(DQI)<br />

Accuracy<br />

Measurement<br />

Per<strong>for</strong>mance<br />

Criteria<br />

No detects ≥ ½<br />

QL<br />

Analyst Accuracy 80%-120%<br />

Analyst Accuracy 75%-125%<br />

Analyst<br />

Analyst<br />

Accuracy<br />

Interferences -<br />

Accuracy/Bias<br />

- Precision<br />

≤10% <strong>for</strong><br />

analytes with<br />

concentration<br />

>50x QL<br />

80%-120%<br />

NA NA NA Interferences NA<br />

Notes:<br />

1<br />

Batch is equivalent to 20 or fewer samples prepared and analyzed together with common QC samples.<br />

Final Sampling and Analysis Plan<br />

Installation Restoration Site 29, Hangar 1<br />

Former Naval Air Station, <strong>Moffett</strong> <strong>Field</strong>, <strong>Moffett</strong> <strong>Field</strong>, Cali<strong>for</strong>nia<br />

DCN: AMEC-8816-0005-0039<br />

April 2010<br />

Page 28-3