Instrument<br />

WORKSHEET #24: ANALYTICAL INSTRUMENT CALIBRATION TABLE<br />

This worksheet provides a list of analytical instrumentation required to per<strong>for</strong>m the laboratory analyses and the associated<br />

calibration procedures. In addition, documentation of the frequency, acceptance criteria, and corrective action requirements are<br />

also provided. All instruments will be calibrated according to the schedule specified by the method and instrument manual or<br />

applicable SOPs. This worksheet will be completed upon laboratory procurement.<br />

Calibration<br />

Procedure<br />

GC ICAL Initially, prior to sample<br />

analysis<br />

GC Second source<br />

initial calibration<br />

verification (ICV)<br />

GC CCV Daily be<strong>for</strong>e sample<br />

analysis, every 10<br />

samples, and at the end of<br />

the analysis sequence<br />

GC/MS<br />

ICAL<br />

Table 24-1. Analytical Instrument Calibration<br />

Frequency of Calibration Acceptance Criteria Corrective Action<br />

One of the options below:<br />

Option 1: relative standard<br />

deviation (RSD) <strong>for</strong> each<br />

analyte ≤ 20%<br />

Option 2: linear least<br />

squares regression:<br />

r≥0.995<br />

Option 3: non-linear<br />

regression: coefficient of<br />

determination (COD) r 2 ≥<br />

0.99.<br />

After every ICAL All analytes within ± 20%<br />

of expected value from the<br />

ICAL.<br />

Method inception, CCAL<br />

fails<br />

All analytes within ± 20%<br />

of expected value from the<br />

ICAL.<br />

Locate the source of the problem. If<br />

expected RSD is not met, then check<br />

<strong>for</strong> standard degradation or per<strong>for</strong>m<br />

instrument adjustment and/or<br />

maintenance to correct the problem and<br />

repeat ICAL.<br />

Prepare fresh standard and re-analyze<br />

ICV to rule out standard degradation or<br />

inaccurate injection. If problem persists,<br />

then per<strong>for</strong>m instrument adjustment<br />

and/or maintenance to correct the<br />

problem and repeat ICAL.<br />

Prepare fresh standard and re-analyze<br />

CCV to rule out standard degradation<br />

or inaccurate injection. If problem<br />

persists, then per<strong>for</strong>m instrument<br />

adjustment and/or maintenance to<br />

correct the problem and repeat ICAL.<br />

Person<br />

Responsible <strong>for</strong><br />

Corrective Action<br />

Laboratory Analyst<br />

Laboratory Analyst<br />

Laboratory Analyst<br />

SOP<br />

Reference<br />

BCORG019<br />

BCORG005<br />

BCORG019<br />

BCORG005<br />

BCORG019<br />

BCORG005<br />

RRF ≥ 0.05, %RSD