Instrument/<br />

Equipment<br />

Mass<br />

Spectrometer<br />

Mass<br />

Spectrometer<br />

Mass<br />

Spectrometer<br />

Mass<br />

Spectrometer<br />

Maintenance<br />

Activity<br />

Pump Oil Level<br />

Detector<br />

Bakeout<br />

Clean source,<br />

replace filament<br />

Detector Precalibration<br />

Testing<br />

Activity<br />

NA<br />

Observation of<br />

deteriorating<br />

chromatograp<br />

hy<br />

Observation of<br />

deteriorating<br />

chromatograp<br />

hy<br />

Resolution<br />

adjustment<br />

Inspection<br />

Activity<br />

Visual<br />

inspection of<br />

oil gauge<br />

and color of<br />

oil<br />

NA<br />

Visual<br />

inspection of<br />

filaments<br />

Observation<br />

of tuning<br />

spectrum<br />

ICP Parameter Setup Physical check Physical<br />

check<br />

Frequency<br />

Check<br />

Monthly;<br />

Change<br />

Annually<br />

As needed<br />

As needed<br />

Every 12<br />

hours<br />

Initially; prior<br />

to DCC<br />

Acceptance Criteria<br />

Adequate oil<br />

level<br />

No interference/<br />

carryover peaks<br />

No interference;<br />

adequate<br />

spectral<br />

resolution<br />

Per method<br />

tuning<br />

requirements<br />

Predetermined<br />

optimum parameter<br />

settings<br />

Corrective<br />

Action<br />

Add oil<br />

Clean<br />

detector<br />

Clean<br />

source;<br />

replace<br />

filament<br />

Adjust<br />

optics<br />

Reset if<br />

incorrect<br />

Responsible<br />

Person<br />

Laboratory<br />

Analyst<br />

Laboratory<br />

Analyst<br />

Laboratory<br />

Analyst<br />

Laboratory<br />

Analyst<br />

Laboratory<br />

Analyst<br />

SOP<br />

Reference<br />

BCORG020<br />

BCORG020<br />

BCORG020<br />

BCORG020<br />

BCMET013<br />

Final Sampling and Analysis Plan<br />

Installation Restoration Site 29, Hangar 1<br />

Former Naval Air Station, <strong>Moffett</strong> <strong>Field</strong>, <strong>Moffett</strong> <strong>Field</strong>, Cali<strong>for</strong>nia<br />

DCN: AMEC-8816-0005-0039<br />

April 2010<br />

Page 25-2