The m<strong>in</strong>imum detectable activity (MDA) characterizes the system before perform<strong>in</strong>g any<br />

measurements, therefore it is def<strong>in</strong>ed on the basis of the detection limit. It is calculated as:<br />

equation 2.3. 13<br />

where:<br />

t M is the time of measurement;<br />

y is the yield, i.e. the probability of emission of the considered radiation per nuclear<br />

transformation;<br />

ε is the efficiency factor.<br />

In case of direct measurements of activity <strong>in</strong> organs or whole body as considered <strong>in</strong><br />

the aforementioned equations, “background” has a particular mean<strong>in</strong>g because it is a “not<br />

contam<strong>in</strong>ated subject ”. However the person’s background is determ<strong>in</strong>ed by the radionuclides<br />

that are naturally present with<strong>in</strong> body and that can significantly change from subject to<br />

subject. The MDA calculated on the basis of its background spectrum will be characteristic<br />

for such person. In order to have a more general description of MDA capability of a detect<strong>in</strong>g<br />

system, a general background spectrum is normally calculated as average of background<br />

spectra of a reference group of not contam<strong>in</strong>ated persons. The evaluations of MDA are then<br />

carried out on this average spectrum.<br />

DIN methodology is not applied to a phoswich detect<strong>in</strong>g system. For such detector the<br />

MDA is calculated by substitut<strong>in</strong>g the L D with 2 times the standard deviation of the net counts<br />

rate <strong>in</strong> the region of <strong>in</strong>terest <strong>in</strong> the spectrum [136]. The spectrum is evaluated by subtraction<br />

of four reference person’s background spectra with similar body proportions and potassium<br />

content. From the result<strong>in</strong>g differences the standard deviation is calculated.<br />

2.3.2 MATERIALS AND METHODS<br />

2.3.2.1 Description of the devices<br />

MDA = L D<br />

t Mεy<br />

The direct measurements of activity <strong>in</strong> organs and tissues were performed with both<br />

HPGe and phoswich detect<strong>in</strong>g systems. Both systems are located <strong>in</strong> a shielded room with<br />

approximately <strong>in</strong>ner dimensions: height 2.25 m; width 2.25 m; lenght 4.20 m. The room is<br />

accessible through a pneumatically moved steel door and shielded with four different layers.<br />

In decreas<strong>in</strong>g order of Z they are: Steel (15 cm), Lead (3 mm), T<strong>in</strong> (1.5 mm) and Copper (0.5<br />

mm). The room is also equipped with a double ventilation system operat<strong>in</strong>g at 292 cm 3 h -1 and