366D000M 1/4/2011SG-HE ONLYSG-HE cabinets utilize the Control Resources Motor Control as seen above. Instructions to adjust motor speedare as follows:I. Adjusting the Control Resources VFD1.) Before powering cabinet, ensure that the jumper selection is wired for the unit’s incoming voltage;top 2 terminals (115V) lower 2 terminals (230V).2.) Verify that the wires secured to the terminal block are in this order from top to bottom, motor white,motor red, motor black, ground green or green/yellow, line white, line black.3.) Once the unit is powered up, standard intake and down flow measurements can be made. In order toadjust the motor controller, the left DIP switch must be in the downward position to switch into PGMmode107