366D000M 1/4/20114.) With the drive in PGM, the blue speed potentiometer can be used to increase (turn clockwise) ordecrease (turn counterclockwise) the cabinet airflow.5.) After the cabinet damper and motor controller are set to produce the desired intake and supplyairflow, allow the drive to operate for at least 1 minute in PGM mode. After 1 minute, set the left DIPswitch to the (AUTO) position. This switches the drive into the standard operating mode.6.) Once in AUTO, the drive will automatically compensate for filter loading, and will maintain setpoint airflow for the life of the HEPA filters.108