General InformationOffline CalibrationThe offline calibration feature of the <strong>2110XP</strong> analyzer allows the operatorto adjust the analyzer to values determined by alternate methods used intheir laboratory such as elemental spectroscopy and ion chromatography.The offline calibration is essentially a one point calibration. To perform anoffline calibration, a sample is taken from the bypass of the analyzer; thesample concentration value is stored in memory; the sample is analyzedby an alternate method of choice; the previously stored reading is adjustedto the lab method result; and the analyzer is then returned to the analysismode. The term “offline calibration” refers only to the fact that a samplefrom <strong>2110XP</strong> analyzer bypass is taken “offline” for laboratory analysis; infact, no downtime is experienced during the procedure and the analyzerremains online throughout.<strong>Thermo</strong> <strong>Scientific</strong> <strong>Orion</strong> <strong>2110XP</strong> <strong>Ammonia</strong> <strong>Analyzer</strong> User GuideI-6