TroubleshootingMalfunction Possible Cause RemedyOver-RangeReadingsElectrode connectionsloose or badElectrode wiringBlocked junction onreference electrode<strong>Ammonia</strong> or referenceelectrode failureElectronics failure orbad setupVerify the electrode connections to theelectrode cables.Electrodes not properly wired to terminalstrip. Refer to Chapter II.Make sure electrolyte is flowing properly.Electrode is shorted, cracked or damaged.Install a new electrode. Refer to Chapter II.Reset the analyzer. Refer to Chapter VI.<strong>Analyzer</strong> out of calibration Recalibrate the analyzer. Refer to Chapter IV.Low Readings <strong>Analyzer</strong> out of calibration Recalibrate the analyzer. Refer to Chapter IV.Sample is very pureElectronics failureCheck sample status. Perform a span checkand if span check passes, the analyzer is okay.Try resetting the analyzer. Refer toChapter VI or consult Technical Support.Default Battery failure Consult Technical Support.Electronics failureReset the analyzer. Refer to Chapter VI.<strong>Thermo</strong> <strong>Scientific</strong> <strong>Orion</strong> <strong>2110XP</strong> <strong>Ammonia</strong> <strong>Analyzer</strong> User GuideVI-13