AppendixOutputsNumber of Analog OutputsAnalog OutputsOutput SelectionsTwo, one dedicated to ammonia, one dedicated totemperature and shared groundGalvanically isolated0 to 20 mA or 4 to 20 mAProgrammable RangeYesChannel Assignments<strong>Ammonia</strong> and temperatureRelative Accuracy ± (0.05 mA + 0.5%)Maximum Load500 ohm or 10 VLog and Linear Output OptionsYes, operator selectableAlarm OutputsNumber of Relay Outputs 3Maximum Relay LoadMinimum Value AlarmMaximum Value AlarmError AlarmCalibration/offline AlarmProgrammable Min. and Max. Values250 VAC/5 A, 30 VDC/5 A, resistive load onlyYesYesYesYesYes<strong>Analyzer</strong> FeaturesNon-volatile MemoryBattery Backup (Replaceable)Data LoggingKeypadYesYesYes8 electromechanical keys with graphic overlaySoftware FeaturesSelf-test and DiagnosticsReal Time ClockPassword ProtectionReset FunctionProgrammable AlarmsMeasurement, ActiveCalibrationYesYesYesYesHigh, low, error, calibration/offlineSignal noise, temperatureSensor response, drift, E 0 , slope, temperature<strong>Thermo</strong> <strong>Scientific</strong> <strong>Orion</strong> <strong>2110XP</strong> <strong>Ammonia</strong> <strong>Analyzer</strong> User GuideA-5