<strong>Analyzer</strong> PreparationMounting andPlumbing InstructionsRefer to the Appendix, Mounting Dimensions section.Warning: Do not connect power prior to the mounting and plumbing ofthe analyzer.Recommendations• Select a site for the analyzer that allows it to be permanently boltedwith ample height for atmospheric drain operation. Be sure that there isready access to the electronic controls, calibration port and electrodes.• A clearance of 15 inches (about 40 cm) must be allowed above theflow cell calibration port. Insert the pipet vertically (not angled) duringthe calibration.• The analyzer location must permit connections to a sample line, drainand AC power supply and any connections for output devices.• The analyzer should be mounted as close to the sampling point aspossible. This ensures the fastest possible response to a changing samplecondition. Refer to the Appendix, Sample Conditions section.• For proper flow cell operation, the analyzer must be installed straightand level upon its mounting location. Failure to level the analyzer maycause poor siphoning in the flow cell.Instructions1. Prepare the mounting holes. Carefully lift the analyzer and bolt it intoplace. Do not lift the analyzer by holding on to any of the plumbing orfluid handling components.2. Connect a waste line to the outlet of the analyzer, which is 3⁄4 NPTmale. The waste line should be connected to a drain of sufficientcapacity, 0.5 inch (1.27 cm) OD is recommended.3. Connect a sample line to the inlet of the analyzer, which is 1⁄4 NPTfemale. It is recommended that a shutoff valve be installed at thesampling point.4. The analyzer must be mounted and leveled vertically for properoperation.<strong>Thermo</strong> <strong>Scientific</strong> <strong>Orion</strong> <strong>2110XP</strong> <strong>Ammonia</strong> <strong>Analyzer</strong> User GuideII-2

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!