CalibrationError Operator Action Scrolling Marquee Main Display NotesCalibration Errors (cont’d)Error110Press to continue the calibrationdespite the bad slope.orPress to abort the calibration andrefer to Exit 3.CH2 BAD SLOPEPRESS ENTERE110The calculated slopewas outside of therecommended range.Perform electrodecleaning andconditioning, refer tothe troubleshootingsection and repeat thecalibration.Error125Press to abort the calibration andrefer to Exit 3.Verify that the correct standards wereused during the calibration, the standardvalues were entered correctly and in theright order and the electrodes are workingproperly and then repeat the calibration.CH1 BAD OFFSETPRESS ENTERE125Use new standards.Perform electrodecleaning, conditioningand troubleshooting.Error126Press to abort the calibration andrefer to Exit 3.Verify that the correct standards wereused during the calibration, the standardvalues were entered correctly and in theright order and the electrodes are workingproperly and then repeat the calibration.CH2 BAD OFFSETPRESS ENTERE126Use new buffers orstandards.Perform electrodecleaning, conditioningand troubleshooting.<strong>Thermo</strong> <strong>Scientific</strong> <strong>Orion</strong> <strong>2110XP</strong> <strong>Ammonia</strong> <strong>Analyzer</strong> User GuideIV-12