Thermo Scientific Orion 2110XP Ammonia Analyzer
Thermo Scientific Orion 2110XP Ammonia Analyzer
Thermo Scientific Orion 2110XP Ammonia Analyzer
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
CalibrationStep Operator Action Scrolling Marquee Main Display Notes17 Wait 0 to 8 minutes. AWAITING DATA STABILITY Actual readingsflashingSystem continuouslyevaluates thedata until a stablecondition is met.If there is a calibrationerror, the appropriateerror message will bedisplayed.18 Add standard 2 through thecalibration port.ADD STD2PRESS ENTERAdd Std219 Press . ADD STD2PRESS ENTER20 Wait 2 minutes. MEASURING STD2CIRCULATINGWAIT 2 MINUTESAdd Std2Actual readingsflashingSystem circulatessample. Minutes willbe counted down onthe marquee.21 Wait 0 to 8 minutes. AWAITING DATA STABILITY Actual readingsflashingSystem continuouslyevaluates thedata until a stablecondition is met.If there is a calibrationerror, the appropriateerror message will bedisplayed.22 No action necessary. System calculatesnew Eo and slope andchecks for errors.If there is a calibrationerror, the appropriateerror message will bedisplayed.<strong>Thermo</strong> <strong>Scientific</strong> <strong>Orion</strong> <strong>2110XP</strong> <strong>Ammonia</strong> <strong>Analyzer</strong> User GuideIV-8