<strong>Analyzer</strong> MaintenanceReplacement ofRestrictor TubingUse the restrictor tube assembly (Cat. No. 2100RT). Refer to Figure V-3for the restrictor tubing connections.1. Close the inlet valve to stop sample flow through the analyzer.2. Loosen the restrictor tube fitting connection from the top of theflow meter.3. Loosen the restrictor tube fitting from the left side of the flowcell block.4. Pull out the restrictor tube assembly at each end.5. Unscrew the clip behind flow cell that holds the restrictor tube tothe panel.6. Replace with a new restrictor tubing assembly.7. Be sure all of the fittings are secure, to prevent leaks.8. Open the inlet valve to start sample flow through the analyzer.Figure V-3Restrictor Tubingrestrictoradapterrestrictortube assemblyfittingclampflowmeterO-ringto check valve onflow cell manifold<strong>Thermo</strong> <strong>Scientific</strong> <strong>Orion</strong> <strong>2110XP</strong> <strong>Ammonia</strong> <strong>Analyzer</strong> User GuideV-6