I-1. Questions & Answers about <strong>Karl</strong> <strong>Fischer</strong> <strong>Reagents</strong><br />
3. Check Solution P<br />
Check Solution P is a solution <strong>of</strong> organic solvent containing a trace amount<br />
<strong>of</strong> water (approximately 4mg/ml). It is used to adjust the end point <strong>of</strong><br />
coulometric titration by providing surplus water when free iodine is produced<br />
in the electrolytes. It can also be used for day-to-day inspections <strong>of</strong><br />
coulometric titration systems. Since methanol is not used as the organic<br />
solvent, this product can also be used with both general-purpose<br />
Aquamicron products and products prepared for use with ketones.<br />
Moisture Content Packaging<br />
3.8-4.2mgH2O/ml 100ml glass bottles with septum caps<br />
[Method <strong>of</strong> Use]<br />
A red tinge in the anolyte solution indicates the presence <strong>of</strong> free iodine.<br />
Normally, the measurement <strong>of</strong> moisture content cannot be initiated. In such<br />
cases, Check Solution P is added to the anolyte as required (until the red<br />
tinge disappears) to supply surplus water and turn all <strong>of</strong> the iodine into<br />
iodide ions. The end point for the anolyte solution can be fine-tuned by<br />
adding Check Solution P in 100ul increments until the red tinge disappears.<br />
Titration is then started.<br />
Check Solution P can also be used for day-to-day verification <strong>of</strong> moisture<br />
meter. This is done by precisely measuring the moisture content <strong>of</strong> the<br />
Check Solution P itself.<br />
4. Water Standard Solution<br />
Aquamicron Water Standard Solution is a water solution containing a<br />
standard amount <strong>of</strong> water. It is used to verify coulometric moisture meters<br />
and to determine the titer <strong>of</strong> KF reagents when used in volumetric moisture<br />
meters. The following products are available.<br />
36<br />
Product Application Packaging<br />
Aquamicron Water Standard 0.2mg<br />
Aquamicron Water Standard 1mg<br />
Aquamicron Water Standard 10mg<br />
Verification <strong>of</strong> moisture meters<br />
(coulometric titration)<br />
Verification <strong>of</strong> moisture meters<br />
(coulometric titration)<br />
Determination <strong>of</strong> titer <strong>of</strong> KF reagents<br />
(volumetric titration)<br />
5ml ampoules (10 per case)<br />
5ml ampoules (10 per case)<br />
8ml ampoules (10 per case)