I-1. Questions & Answers about <strong>Karl</strong> <strong>Fischer</strong> <strong>Reagents</strong><br />

3. Check Solution P<br />

Check Solution P is a solution <strong>of</strong> organic solvent containing a trace amount<br />

<strong>of</strong> water (approximately 4mg/ml). It is used to adjust the end point <strong>of</strong><br />

coulometric titration by providing surplus water when free iodine is produced<br />

in the electrolytes. It can also be used for day-to-day inspections <strong>of</strong><br />

coulometric titration systems. Since methanol is not used as the organic<br />

solvent, this product can also be used with both general-purpose<br />

Aquamicron products and products prepared for use with ketones.<br />

Moisture Content Packaging<br />

3.8-4.2mgH2O/ml 100ml glass bottles with septum caps<br />

[Method <strong>of</strong> Use]<br />

A red tinge in the anolyte solution indicates the presence <strong>of</strong> free iodine.<br />

Normally, the measurement <strong>of</strong> moisture content cannot be initiated. In such<br />

cases, Check Solution P is added to the anolyte as required (until the red<br />

tinge disappears) to supply surplus water and turn all <strong>of</strong> the iodine into<br />

iodide ions. The end point for the anolyte solution can be fine-tuned by<br />

adding Check Solution P in 100ul increments until the red tinge disappears.<br />

Titration is then started.<br />

Check Solution P can also be used for day-to-day verification <strong>of</strong> moisture<br />

meter. This is done by precisely measuring the moisture content <strong>of</strong> the<br />

Check Solution P itself.<br />

4. Water Standard Solution<br />

Aquamicron Water Standard Solution is a water solution containing a<br />

standard amount <strong>of</strong> water. It is used to verify coulometric moisture meters<br />

and to determine the titer <strong>of</strong> KF reagents when used in volumetric moisture<br />

meters. The following products are available.<br />

36<br />

Product Application Packaging<br />

Aquamicron Water Standard 0.2mg<br />

Aquamicron Water Standard 1mg<br />

Aquamicron Water Standard 10mg<br />

Verification <strong>of</strong> moisture meters<br />

(coulometric titration)<br />

Verification <strong>of</strong> moisture meters<br />

(coulometric titration)<br />

Determination <strong>of</strong> titer <strong>of</strong> KF reagents<br />

(volumetric titration)<br />

5ml ampoules (10 per case)<br />

5ml ampoules (10 per case)<br />

8ml ampoules (10 per case)