30.11.2012 Views

Development of Karl Fischer Reagents

Development of Karl Fischer Reagents

Development of Karl Fischer Reagents

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

Silicon rubber<br />

Figure 6: Sampling Syringe<br />

Electrolysis cell<br />

Chapter I: Basic Knowledge<br />

Figure 7: Sample Injection System<br />

Catholyte chamber<br />

Anolyte chamber<br />

5. Moisture Vaporization<br />

With this method, the coulometric titration system is used in conjunction with<br />

a moisture vaporization system. Approximately 150ml <strong>of</strong> anolyte<br />

(Aquamicron AX or AS) is placed in an electrolysis cell. Two ampoules<br />

(10ml) <strong>of</strong> catholyte (Aquamicron CXU) are used.<br />

The condition for moisture vaporization should be optimized for the<br />

particular sample. Detailed instructions for various types <strong>of</strong> samples are<br />

provided later in this manual.<br />

53

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!