II-1 Organic Compounds<br />

(2) Coulometric titration<br />

<strong>Reagents</strong> used: Aquamicron AKX 100ml<br />

Aquamicron CXU 5ml<br />

84<br />

Substance<br />

Acetone<br />

Methyl ethyl ketone<br />

Methyl isobutyl ketone<br />

Acetylacetone<br />

Acetophenone<br />

Cyclopentanone<br />

Cyclohexanone<br />

Diacetone alcohol<br />

2-Methoxy-2-methyl-4 -pentanone<br />

Sample quantity<br />

(g)<br />

0.3954<br />

0.4141<br />

0.8010<br />

0.9760<br />

0.5451<br />

0.4955<br />

0.3117<br />

0.9602<br />

1.8963<br />

Measurement value<br />

(µg)<br />

907<br />

464<br />

102<br />

683<br />

215<br />

1090<br />

635<br />

217<br />

81<br />

MoistureContent<br />

(%)<br />

0.229<br />

0.112<br />

127ppm<br />

0.700<br />

394ppm<br />

0.220<br />

0.204<br />

226ppm<br />

043ppm<br />

Measurable<br />

amount (ml)<br />

It is important to be aware <strong>of</strong> patterns <strong>of</strong> blank current (also described as<br />

"background" or "drift" current) in the moisture meter. If a gradual rise in the<br />

blank current causes the end point to fluctuate, it will not be possible to<br />

obtain accurate analytical results. In such cases the electrolyte must be<br />

replaced.<br />

10<br />

15<br />

15<br />

15<br />

15<br />

5<br />