<strong>XNX</strong> <strong>Universal</strong> <strong>Transmitter</strong>FaultNumber Description Contents Of Data Field Condition RecoveryF141End of Cell LifeEC EC CartridgemV mV Personality BoardIR N/AStabilization TimeoutDiagnostic DataInstalled sensor exceeded sensor lifeparameterFit replacement cartridge.F143F145Diagnostic DatamVIRSensor Exceeded Expected StabilizationTimeEC Unstable Sensor OutputReflex FailureEC EC Cartridge Diagnostic DatamVIR N/A N/AGeneral Optical FaultSensor exceeds normal warm-up timeEC cell has reached end of life.Cycle power, contact HA Service if problempersists.Fit replacement cell or cartridge.F146F148F149F150ECmV N/A N/AIR Excel/Optima IR Sensor Fault CodeOption Board FailureECmV All PersonalitiesDiagnostic DataIRInternal Communication Failure (mA)ECmV All PersonalitiesN/AIRmA Output Monitoring FailECmV All PersonalitiesActual measured mA output valueIRInternal option board hardware failure.Internal 4-20 mA monitoring circuitcommunication failure.mA not producing expected levels.Contact HA ServiceContact HA ServiceContact HA ServiceContact HA ServiceSection 5 - Warnings/Faults 5-105.4 Fault Messages