PC MaintenanceThe initial Gas Calibrati<strong>on</strong> window opens.Figure 3G-1: Initial Gas Calibrati<strong>on</strong> Screen✓ A warning displays and will automatically disappear followed by an autozero.2. The system then performs a zero calibrati<strong>on</strong>. In zero calibrati<strong>on</strong>, the gasanalyzer submits clean air, room air, into the sample system to verify thatthe analyzer reads properly when there is no gas present in the sample cell.✓ Make sure the vehicle exhaust sample probe is in fresh air, NOTsampling vehicle exhaust of any type.3. The Calibrati<strong>on</strong> Bottle Values screen then displays. Verify or change thevalues to match your calibrati<strong>on</strong> gas.— To change values, Tab to the value you want to change, press Deleteand input the correct value.Figure 3G-2: Calibrati<strong>on</strong> Bottle Values Screen4. Press OK. The Attach Calibrati<strong>on</strong> Bottle screen displays.Figure 3G-3: Attach Calibrati<strong>on</strong> Bottle Screen3G-2