Pocket PC MaintenanceIf the gas calibrati<strong>on</strong> fails, the Gas Calibrati<strong>on</strong> Failed screen displays.Figure 2G-9: Gas Calibrati<strong>on</strong> Failed Screen1. Verify that the gas bottle was open.2. Make sure pressure is present at the calibrati<strong>on</strong> gas hose outlet. (Gasmust be coming out the end of the hose.)3. Verify that the gas bottle was properly plugged into the gas bench.4. Verify that the gauge <strong>on</strong> the calibrati<strong>on</strong> gas regulator read at least 30 psi.5. Verify the proper calibrati<strong>on</strong> gas bottle values.6. Rerun the calibrati<strong>on</strong>.Resetting NO x Cell Output Voltage: From time to time during normal use andcalibrati<strong>on</strong>, the NO x cell output voltage, used to determine NO x values, must bereset. If after a gas calibrati<strong>on</strong> the NO x Warning screen shown in Figure 2G-10appears, perform the gas calibrati<strong>on</strong> again. However, DO NOT use the Calibrati<strong>on</strong>functi<strong>on</strong>. Instead select and perform the New NO x Sensor sequence (Maintenance> Setup > New NO x ); this procedure resets the NO x cell output voltage. (See NewNOx Sensor, page 2G-18.)Figure 2G-10: NO x Warning Screen2G-6